7. Lower the front edge of the diffuser until clears the bottom of the sash, and then

pull the diffuser straight out of the cabinet, as shown in Figure 5-4.

Figure 5-4
















8. The Logic is now ready for the HEPA Filter Leak Test. Set the photometer

sampling valve to “DOWNSTREAM”. Ensure there is a proper vacuum at the
sampling port of the pistol.

9. Open the following number of Laskin nozzle(s):

For all 3-foot Logic models – 1 nozzle

For all 4-, 5- and 6-foot Logic models – 2 nozzles

10. Ensure the pressure gauge on the generator is reading 20 +/- 1 PSIG.

11. Scan the downstream side of the HEPA filter by passing the sampling nozzle of

the gun in slightly overlapping strokes over the entire surface of both of the
filters, with the sampling port not more than 1 inch from the surface of the filter
media. Scan the entire periphery of the filters and the gaskets between the filter
frame and the Logic frame. Scanning shall be done at a traverse rate of not more
than 2 inches per second.


Note: Operation of the photometer may become erratic when testing the front edge of
the supply filter frame, due to aspiration of room air into the front of the work area. This
problem can be minimized or eliminated by placing the edge of a sheet of rigid plastic or
metal at the edge of the filter gasket when scanning this area, as shown in Figure 5-5.

Figure 5-5