Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 157

Advertising
background image

Page 157 of 197

Section 2. - Canopy Pressure Switch Wiring Harness
Installation Procedure

1. Ensure that the Logic is unplugged.
2. Unwind the Switch wiring harness from the alarm module on the right side of the

canopy.

3. Connect the Harness to the Alarm Contact connector on the left side of the

electronics module, as shown. Ensure that there is slack in the sensor cable
where it connects to the harness.


NOTE: The female blade connectors on the harness are used for signaling external
systems the blower and alarm status. Refer to the unit wiring diagram for further
information.




NOTE: Ensure the sensor harness
does not contact any moving
parts.
Wiring Harness



Section 3. – Control Board Installation Procedure

1. Ensure that the Logic is unplugged.
2. Open the right side panel of the Logic by removing the two Phillips screws on the

front edge of the panel.

3. Let the panel swing out, and examine the control board on the side of the

cabinet as shown.

4. Note the orientation of the ribbon and touchpad connectors before disconnecting

them from the board.

5. Loosen the bottom nut, and remove

the top nut to remove the control
circuit board assembly from the
chassis of the Logic