3. Move the base stand into its final location. Using a carpenter’s level, adjust each
leveling foot until the stand is level in both planes as shown in Figure 3-4. You are now
ready to lift the Purifier onto its stand.
Figure 3-4
Manual or Electric Hydraulic Lift Base Stands
These base stands offer infinitely adjustable height between 25.5 and 33.5 inches,
giving a work surface height of 28.0 to 36.0 inches. The height is adjusted either by a
manual (hand crank) or electric pump that drives hydraulic rams in the legs of the
stands. All of the hydraulic stands are equipped with fixed feet, but can be converted to
casters with the addition of Caster Wheel Kit #3784000. The Base Stands for each
Logic model is listed in Table 3-3 below.
Table 3-3
Width
Manual Lift
Stand #
Electric (115V) Lift
Stand #
Electric (230V) Lift
Stand #
3' 3780200
3780100
3780103
4' 3780201
3780101
3780104
5' 3780203
3780106
3780107
6' 3780202
3780102
3780105
Note: When installing the hydraulic lift base stand, ensure that the hydraulic lines and
the electrical cord are clear of any obstructions before installing the cabinet on the stand
or operating the lift system.