Figure 8-13
Center Sash Support
3. Remove the last two screws. Gently rock the cover, slide it to the left or
right, and swing it out and away from the unit.
4. Use a 1/2-inch socket or wrench to tighten the three plenum lift bolts located
as noted in Figure 8-14. This will lift the plenum off of the supply HEPA filter.
Remove the Supply HEPA filter retainers on the front of the plenum, if
present.
5. Slide the supply HEPA filter straight out the unit.
6. With the HEPA filters removed, inspect the filter shelf for damage.
Figure 8-14
Plenum Lift Bolts (3)
Supply HEPA filter retainers (3)
To Install a Supply HEPA Filter:
1. Install a new supply HEPA filter by pushing it straight into the cabinet,
ensuring that it is correctly oriented, and that the filter is centered on the
filter shelf.
2. Use a 1/2-inch socket or wrench to sequentially loosen the three plenum lift
bolts located as noted in Figure 8-14. When properly positioned, the plenum
box will be aligned with the supply HEPA filter frame. Continue to loosen the