Vibration (optional)

The vibration level should be measured as described in ANSI/NSF Standard 49:2007.

Note: The vibration meter should have a range of 20-20,000 Hz, and have an
output in displacement.

1. Vibration level should be recorded from a single point, on the geometric center

of the work surface.

2. Take a background reading with the cabinet blower off at the test point.

3. Start the cabinet blower, and take a vibration reading.

Acceptance

Vibration should be less than 0.0002 inches displacement when the cabinet is new.


UV intensity (optional)

Note: The UV radiometer should measure light at a wavelength of 254nm.

1. UV intensity should be recorded from a single point, on the geometric center of

the work surface.

2. Take a background reading with the cabinet blower off at the test point.

3. Close the sash, turn on the UV light, and let it warm up for at least 5 minutes.

Acceptance

The UV intensity should be at least 200 uW-cm

2

.


Most common causes for insufficient UV light levels are:

a. Malfunctioning lamp(s)

b. Old lamp(s)

c. Dusty/dirty lamp surfaces