Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 176
![background image](/manuals/738628/176/background.png)
Page 176 of 197
Postfilter Conversion Kit No. 3859300, -01, -
02, -03
NOTE: To prevent damage to the Exhaust HEPA filter, exercise caution when
removing or installing the exhaust filter cover or the postfilter pan.
Installation Procedure
Note: exercise caution when removing or installing the exhaust filter
cover, to prevent damaging the Exhaust HEPA filter.
On Models Manufactured prior to April 2009 –
8. Turn off the Biosafety Cabinet.
9. Remove the exhaust HEPA filter cover by lifting and compressing the
cover until its tabs are clear of the exhaust filter bracket, as shown.
Lift the center of the cover up while
compressing the cover.
Cover tabs in bracket slots
10. This type cover design is not compatible with the postfilter accessory.
Discard this cover and contact Labconco’s Product Service department at
1-800-821-5525 for a replacement redesigned exhaust filter cover.