Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 21
Page 21 of 197
QuickChart - Logic Type B2 Biohazard Cabinets (cont)
Exhaust HEPA Filter Leak Test Data
Air Displacement (CFM)
723
1083
Laskin Nozzles needed
2
2
Theoretical aerosol conc. (ug/l)
1
38
25
Actual aerosol conc. (ug/l)
2
21
17
Exhaust HEPA Data
Width (inches)
26
48
Depth (inches)
18
18
Height (inches)
10
5.88
5.88
Performance (CFM)
353
529
Performance (Pressure in. H
2
O) .23-.37
.23-.37
Labconco P/N
3838501
3838503
Filtration Group P/N
42547L
41727L
AAF P/N
907-543-
206
907-543-
208
Motor/Blower Data
Labconco P/N
11
3832204
3832205
Motor HP
½
½
Blower Size (inches)
10 x 8
10 x 8
Nominal Speed Control Setting
74
73
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8
TL741
F40T8
SP41
Fluor. Lamp Labconco Part #
9721900
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
Fluor. Ballast Robertson Part #
ISL332T8
MVW
ISL332T8
MVW
Circuit Breaker Data 100V/115V
Motor/Light Breaker Rating, Amps
8
10
Motor/Light Breaker, Labconco Part #
1327208
1327201
Motor/Light Breaker, Mechanical Products Part #
1600-062-
080-007
1600-062-
100-007
Outlet Breaker Rating, Amps
5
5
Outlet Breaker, Labconco Part #
1327204
1327204
Outlet Breaker, Mechanical Products Part #
1600-062-
050-007
1600-062-
050-007
Circuit Breaker Data 230V
Motor/Light Breaker Rating, Amps
4
5
Motor/Light Breaker, Labconco Part #
1327209
1327204
Motor/Light Breaker, Mechanical Products Part #
1600-062-
040-007
1600-062-
050-007
Outlet Breaker Rating, Amps
3
3
Outlet Breaker, Labconco Part #
1327200
1327200
Outlet Breaker, Mechanical Products Part #
1600-062-
030-007
1600-062-
030-007