Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 164

Advertising
background image

Page 164 of 197

To Reset the Filter Gauge


To reset the filter gauge to display 100%, select the “reset filter gauge” option,

then press:

“Mute/OK”

To leave the gauge at its current setting, select the option “Filter not changed”,

and press:

“Mute/OK”

The display will now change to the Inflow and downflow calibration mode.
Setting the Inflow and Downflow
The next screen will be the downflow calibration screen.
Press the up or down (▲

or

▼) arrows to increase or decrease the downflow

value respectively until the proper value is displayed. When the number is

properly set press:
“Mute/OK”

The next screen will be the inflow calibration screen. Press the up or down

(▲

or

▼)

arrows to increase or decrease the inflow value respectively until the proper value is
displayed. When the number is properly set press:

“Mute/OK”
This will end the calibration mode, and the unit will automatically turn off and

then restart, displaying inflow and downflow.