Electrical System Service Operations


NOTE: In order to properly service the Logic, you must obtain a Logic Certifier
Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by
contacting Labconco Customer Service Department at 800-821-5525
.

Fluorescent Lamp

Lamp Identification


All Logics use dual fluorescent lamps. The lamp information is shown in Table
8-1. The code of the lamp part number is as follows:

F25T8 TL741
F = fluorescent lamp
25 = the lamp Wattage
T8 = the lamp mounting configuration
TL741 = the phosphor used in the lamp

Note: The phosphor of the lamp determines the color and intensity of
the lamp. Not matching the phosphor may affect the quality and/or
color of the light.

Table 9-1

Fluorescent Lamp Specifications

Model

34308 34300 34408 34400 34508 34500 34608 34600

Fluorescent Lamps (2 each)

F25T8
TL741

F25T8
TL741

F32T8
TL741

F32T8
TL741

F40T8

SP41

F40T8

SP41

F40T8

SP41

F40T8

SP41

Fluor. Lamp Labconco Part #

9721901

9721901

9721900

9721900 9721903 9721903

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

3838100

3838100

3838100

3838100

Fluor. Ballast Robertson Part #

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

Changing the Fluorescent Lamp


1. Unplug the cabinet or turn the System Reset Switch, located on the top of the

cabinet, off.

2. Remove the front dress panel as noted in Figure 7-1.
3. Remove the fluorescent lamps by pulling the lamp sockets straight off each end of

the lamp, and releasing both lamps from the spring clips that secure them in place.

4. Install the new lamps by reversing the removal procedure.

Advertising