Grille Removal


Note:
The grille must be thoroughly decontaminated before removing it.

1. Remove the work surface as described earlier.
2. At one end of the grille, grip the front of grille with one hand, and the back with

the other hand. Pivot that end of the grille upward and inward, paralleling the
angle of the sash, as shown in Figure 7-26.

3. Pull the other end of the sash up and away from the bottom edge of the cabinet.
4. Reinstall the grille by reversing the above sequence, ensuring that the grille

properly engages the bottom edge of the cabinet.

Figure 7-26










Towel Catch Removal

Although not normally required, the towel catch can be removed for cleaning,
inspection, etc.
Note: The work surface of the cabinet and the towel catch must be thoroughly
decontaminated before removing either.
1. Remove the work surface as described above.
2. Lift the front end of the center support straight up to release it from the locating

pins (A). Set the support aside in the work area.

3. Remove the towel catch by pivoting the bottom out toward you (B).
4. Surface decontaminate the towel catch before removing it.
5. Reinstall the towel catch by sliding it back into position, ensuring that the two slots

on either end of the towel catch properly engage the rails on the rear of the duct, as
shown in Figure 7-27. Also ensure that the DOP Sampling Line (the clear tube with a
black cap) either passes through a hole in the towel catch, or passes along the
outside edge of the catch without being kinked.