Preparing the Logic for Gaseous Sterilization

Note: This section only reviews the steps required for preparing the Logic for
Gas Sterilization. Thoroughly understand the sterilization procedures and
protocols supplied by the manufacturer of the sterilizing system before
attempting this operation.
Note: The outlets of the Logic are designed to be continuously energized.

1. Unplug the Logic from electrical power.
2. Thoroughly surface decontaminate the working area of the cabinet.
In order to get a complete seal on the front of the cabinet, do the following:
3. Remove the prefilter on the top of the cabinet and seal the inlet opening with

plastic film and duct tape.

4. Remove the work surface after it has been surface decontaminated.
5. Remove the front grille.
6. Place the work surface and the grille into the work area for decontamination.
7. Remove the front dress panel.
8. Remove the sash stop bracket, located near the top of the blower plenum cover.
9. Lift the sash up until the bottom edge of the sash clears the bottom edge of the

Supply HEPA Filter shelf.

10. Using flexible plastic sheet 5-8 mil thick and duct tape seal the perimeter of the

front opening.

11. Remove the exhaust filter cover.
Note: DO NOT attempt to remove the brackets that secure the exhaust filter cover;
they are needed to maintain rigidity on the top of the chassis.
1. Using flexible plastic sheet and duct tape seal the Exhaust filter opening.
2. The Logic is now prepared for gaseous decontamination.