Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 183

Advertising
background image

Page 183 of 197


Relay Contact #2……the “airflow error” contact
This set of contacts (dry contacts pins 1 & 2 of J4 on the electrical enclosure, and the
ORANGE wires on Labconco harness #38468-00) will indicate when the BSC’s has an
airflow error, as described by the conditions below:

Type of Logic

unit

Alarm Condition

LCD display

message

Airflow Error

dry contacts

Comments

All Labconco
Logic Cabinets

Sash out of
position

SASH TOO HIGH No Response

Motor RPM has

shifted abruptly

AIRFLOW ALERT Contacts shorted

HEPA

Filter

approaching end

of life—20%

CHECK FILTER

No response

HEPA

Filter

approaching end

of life-10%

CHECK FILTER
SOON

Contacts shorted

HEPA Filter end

of life- 0%

REPLACE

FILTERS

Contacts shorted

**Note all
specialty Logic

BSC’s below will

have all the

alarm conditions

listed above, plus

specific alarms
listed below”

A2 BSC with

canopy sensor**

( see above)

Canopy Airflow

incorrect

CANOPY ERROR

Contacts shorted

A2 BSC with

airflow sensor**(

see above)

Downflow Airflow

incorrect

DOWNFLOW

ERROR

Contacts shorted

Airflow

sensor

not working

AIRFLOW

SENSOR ERROR

Contacts shorted

B2 BSC**( see

above)

Exhaust airflow

not correct

EXHAUST ERROR Contacts shorted

Airflow

sensor

not working

AIRFLOW

SENSOR ERROR

Contacts shorted