Changing the UV Lamp

Note: For optimum performance, the UV lamp should be changed on an annual
basis.
Caution: The UV lamp and the work area of the cabinet must be thoroughly
surface decontaminated before removing the lamp.
1. Unplug the Logic.
2. Raise the sash to its full open position.
3. Surface decontaminate the UV lamp and the work area of the cabinet.
4. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and

out of its sockets.

5. Install new lamp by reversing the removal procedure.

UV Lamp Socket Replacement


Note: The UV lamp sockets are left- and right-handed. They are not
interchangeable. The Part number for the right socket assembly is
3742000, and the left is part number 3742001.

1. Unplug the Logic

2. Raise the sash to its full open position.

3. Surface decontaminate the work area and the UV lamp.
4. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and

out of its sockets.

5. To replace a socket, Use a #1 Phillips screwdriver to remove the screw.

6. Pull the socket assembly out of the rear wall.
7. Disconnect the wire connectors, and handle the old socket as contaminated

waste.

8. Install the new socket.

9. Reinstall the Phillips screw through the rear wall and capture the lamp socket

bracket. Tighten the screw.

10. Reassemble the unit.