Restoring software default settings, Calibration, Diagnostic test operation – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 61
Page 61 of 197
Restoring Software Default Settings
To restore the default settings to their original configuration:
1. Access the Menu, and select “Cabinet Set Up”. Press “Mute/OK”
2. In the Cabinet Set Up submenu, select “Restore Defaults”. Press “Mute/OK”
3. Press “Mute/OK” again to restore the software defaults. Pressing “Menu” will
escape this submenu without restoring the defaults.
Table of Default Settings
UV Lamp:
Smart-Start, Continuous, Reset UV Lamp
Hourmeter, Set UV Lamp Life to 6000
hours.
Fluorescent Lamp:
Smart-Start, Reset Lamp Hourmeter, Set
Fluorescent Lamp Life to 6000 hours.
Blower: Smart-Start
Units of Measure:
Ft / Min
Security Lock:
Non-secure
Data Output Rate:
Every second
Calibration
This selection is for the use of qualified certifiers only to adjust the blower speed,
the display of the filter life remaining gauge, and the optional inflow and downflow
displays. Refer to Section 5, “Accessing the Certifier’s Menu”.
Diagnostic Test Operation
This selection is for the use of qualified certifiers only to check proper operation of
the keypad, optional airflow and optional canopy flow alarm sensors. For further
information, refer to the Diagnostics Section in Section Nine – Electrical System
Service Operations.