Restoring software default settings, Calibration, Diagnostic test operation – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 61

Advertising
background image

Page 61 of 197

Restoring Software Default Settings

To restore the default settings to their original configuration:

1. Access the Menu, and select “Cabinet Set Up”. Press “Mute/OK”
2. In the Cabinet Set Up submenu, select “Restore Defaults”. Press “Mute/OK”

3. Press “Mute/OK” again to restore the software defaults. Pressing “Menu” will

escape this submenu without restoring the defaults.


Table of Default Settings

UV Lamp:

Smart-Start, Continuous, Reset UV Lamp
Hourmeter, Set UV Lamp Life to 6000
hours.

Fluorescent Lamp:

Smart-Start, Reset Lamp Hourmeter, Set
Fluorescent Lamp Life to 6000 hours.

Blower: Smart-Start

Units of Measure:

Ft / Min

Security Lock:

Non-secure

Data Output Rate:

Every second


Calibration

This selection is for the use of qualified certifiers only to adjust the blower speed,
the display of the filter life remaining gauge, and the optional inflow and downflow
displays. Refer to Section 5, “Accessing the Certifier’s Menu”.

Diagnostic Test Operation

This selection is for the use of qualified certifiers only to check proper operation of
the keypad, optional airflow and optional canopy flow alarm sensors. For further
information, refer to the Diagnostics Section in Section Nine – Electrical System
Service Operations.