Ground fault circuit interrupter (gfci) test – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 82

Advertising
background image

Page 82 of 197

Ground Fault Circuit Interrupter (GFCI) Test

(115/100 Volt models only)

Note: The GFCI tester must be able to simulate a fault of 3mA.

1. Place the tester into any of the Logic’s outlets. The indicator lights of the tester

should indicate normal function, and the GFCI’s indicator LED should be lit.

2. Press the test button of the tester. The indicator lights should now indicate the

circuit is inactive, and the GFCI indicator LED should be off.

3. Reset the GFCI by pressing the RESET button on the right outlet. The tester

should again indicate normal operation, and the GFCI indicator LED will
illuminate.


Figure 5-20


GFCI Indicator LED

GFCI Reset Button

GFCI Test Button




Acceptance

The GFCI trips during the simulated failure, and the indicator LED turns off. Pressing
the Reset button returns the GFCI to normal function.