Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 190

Advertising
background image

Page 190 of 197

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower runs
slower than
normal; speed
control has little
effect on blower
speed. Blower may
take more than 5
seconds to start.

Motor signal wires
damaged or
disconnected

Measure motor output signal
voltage. Inspect signal wiring and
connections

Power

Supply

Board
misconnected or
defective

Test output voltages on the Power
Supply board.

Motor jumper wire
is damaged/
disconnected
(115/100 Volt
models only)

Inspect grey wire #31 on 15 pin
connector for continuity.

Blower motor is
defective

Replace blower motor.

Keypad

disconnected or
defective

Run keypad diagnostics and check
connections.

Fluorescent light
not working

Sash is closed

Open sash – Fluorescent lights will
not work with the sash closed.

Lamp(s)

are

defective

Replace defective lamp(s)

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballasts

Replace lamp ballasts.

Keypad

disconnected or
defective

Run keypad diagnostics and check
connections.