Type b2 – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 14
![background image](/manuals/738628/14/background.png)
Page 14 of 197
ADJUSTING THE BLOWER SPEED, FILTER GAUGE,
AIRFLOW SENSOR AND ALARM SETPOINT
Type B2
* This screen unavailable on units with firmware prior to 10443 Rev. G.
Press the or keys to adjust the blower
speed. The word ‘wait’ will flash until the
blower speed stabilizes. Press ‘OK’ to
accept the new setting.
CERTIFIER USE ONLY
Calibration mode:
Motor Speed Fac tor:
wait
900
950
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST
80
CERTIFIER USE ONLY
Calibration mode:
Filter setup:
1. New Filter
2. Filter Not Changed
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST
Press the or keys to select. Option # 1 will
reset the filter life remaining gauge to 100%.
Option # 2 will leave the gauge unchanged.
Press ‘OK’ to accept your choice.
CERTIFIER USE ONLY
Calibration mode:
Inflow:
FPM
UV to autoset
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST
105
Press the or keys to adjust the
inflow to match the certified value.
Press ‘UV Light’ to reset the inflow to
105 FPM. Press ‘OK’ to accept.
Close the exhaust damper until exhaust
volume equals 80% of nominal volume.
Press ‘OK’ to accept the alarm volume.
Reopen the damper to its normal position.
CERTIFIER USE ONLY
Calibration mode:
reduce inflow to
alarm level,
OK TO SET
MENU TO SKIP
*
The Logic will automatically
turn off and reset itself into
normal operation.