QuickChart – PuriCare Procedure Stations (cont)

Exhaust HEPA Data

34800

34802

34512

34810

34812

Width

(inches)

26 26 36 48 48

Depth (inches)

18

18

18

18

18

Height (inches)

6

5.88

5.88

5.88

5.88

5.88

Performance

(CFM)

353 353 441 529 529

Performance (Pressure in. H

2

O)

.23-.37 .23-.37 .23-.37 .23-.37 .23-.37

Labconco

P/N

3838501 3838501 3838502 3838503 3838503

Filtration Group P/N

42547L

42547L

42673L 41727L 41727L

AAF P/N

907-543-

206

907-543-

206

907-543-

207

907-543-

208

907-543-

208

Motor/Blower Data

Labconco P/N

7

3832201

3832201

3832203

3832203

3832203

Motor HP

½

½

¾ ¾ ¾

Blower Size (inches)

10 x 8

10 x 8

10 x 8

10 x 8

10 x 8

Nominal

Speed

Control

Setting

85 85 85 81 81

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F32T8
TL741

F32T8
TL741

F40T8

SP41

F40T8

SP41

F40T8

SP41

Fluor. Lamp Labconco Part #

9721900

9721900

9721903

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

3838100

Fluor. Ballast Robertson Part #

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

Circuit Breaker Data

100V/115V

Motor/Light Breaker Rating, Amps

8 8 10 10 10

Motor/Light Breaker, Labconco #

1327208

1327208

1327201

1327201

1327201

Motor/Light Breaker,

Mechanical Products Part #

1600-062-

080-007

1600-062-

080-007

1600-062-

100-007

1600-062-

100-007

1600-062-

100-007

Outlet

Breaker

Rating,

Amps

5 5 5 5 5

Outlet Breaker, Labconco #

1327204

1327204

1327204

1327204

1327204

Outlet Breaker, Mechanical

Products Part #

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

Circuit Breaker Data 230V

Motor/Light Breaker Rating, Amps

4 4 5 5 5

Motor/Light Breaker, Labconco #

1327209

1327209

1327204

1327204

1327204

Motor/Light Breaker,

Mechanical Products Part #

1600-062-

040-007

1600-062-

040-007

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

Outlet

Breaker

Rating,

Amps

3 3 3 3 3

Outlet Breaker, Labconco #

1327200

1327200

1327200

1327200

1327200

Outlet Breaker, Mechanical

Products Part #

1600-062-

030-007

1600-062-

030-007

1600-062-

030-007

1600-062-

030-007

1600-062-

030-007

Canopy Data

Labconco Canopy P/N

3858701

3858701

3858702 3858703

3858703

Canopy Alarm Setpoint Vol. (CFM)

8

337-370 323-364 403-454 503-554 483-544

Nominal

Canopy

Vol.

(CFM) 424 413 558 687 668

Canopy

Vol.

Range

(CFM) 404-444 388-437 524-590 654-720 628-707

Canopy Data 2

nd

Generation

9

Labconco Canopy P/N

3898001

3898001

3898002

3898003

3898003

Canopy Slot Area (F

2

)

0.186 0.186 0.236 0.304 0.304

Nominal Canopy Slot Velocity

(FPM)

10

400 400 315 330 330

Nominal Canopy Slot Volume (CFM)

75 75 75 100 100

Canopy Slot Velocity

Alarm Setpoint (FPM)

10

100 100 100 100 100

Total

Exhaust

Volume

(CFM) 412-445 398-439 478-529 603-654 583-644

Canopy Vacuum (In. H

2

O) .15-.25

.15-.25

.15-.25

.15-.25 .15-.25

Advertising