The logic information center layout, Status area – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 47

Advertising
background image

Page 47 of 197

The Logic Information Center Layout

The Purifier's Information Center is an LCD display located on the right side wall at eye
level. The center’s display is divided into three separate areas, as shown in Figure 4-2,
below.

Figure 4-2


Status Area

Data Area

Icon Area



Status Area

The Status area displays the system status during operation. This line can display any of
the following messages:
OK

The motor is operating properly, and the sash is not open too far.


Sash is too high

The sash is open too far for proper operation.


Airflow Alert

The airflow patterns of the cabinet have been changed, resulting in a sudden

change in the motor speed. This is most likely due to a blockage of the grille or the
exhaust filter outlet. It may also be caused by removal of the work surface while the
cabinet is in operation.

System Error

The motor and display circuit board are not communicating properly. DO NOT USE

THE CABINET UNTIL THE PROBLEM HAS BEEN CORRECTED.