Exhaust Requirements

If the customer intends to connect the Purifier to the optional canopy connection kit,
first examine the location to ensure that it is compatible with the cabinet’s exhaust duct.
The area directly above the cabinet’s exhaust port should be clear of structural
elements, water and utility lines, or other fixed obstructions. There should be enough
clearance to allow for the passage of a 10" diameter duct. Avoid cabinet locations that
require either an elbow directly on top of the cabinet’s exhaust connection or an
excessive number of elbows to clear other items. For a further discussion of the
cabinet’s exhaust system requirements, please go to Section 9: Accessories

.

Electrical Requirements

Table 3-1

Model #

Requirements

3430000-09

115 VAC, 60 Hz, 12 Amps

3430800-09

115 VAC, 60 Hz, 12 Amps

3430810-19

100 VAC, 50/60 Hz, 12 Amps

3430820-29

230 VAC, 50/60 Hz, 6 Amps

3440000-09

115 VAC, 60 Hz, 12 Amps

3440800-09

115 VAC, 60 Hz, 12 Amps

3440810-19

100 VAC, 50/60 Hz, 12 Amps

3440820-29

230 VAC, 50 Hz, 6 Amps

3450000-09

115 VAC, 60 Hz, 16 Amps

3450800-09

115 VAC, 60 Hz, 16 Amps

3450810-19

100 VAC, 60 Hz, 16 Amps

3450820-29

230 VAC, 50/60 Hz, 8 Amps

3460000-09

115 VAC, 60 Hz, 16 Amps

3460800-09

115 VAC, 60 Hz, 16 Amps

3460810-19

100 VAC, 50/60 Hz, 16 Amps

3460820-29

230 VAC, 50/60 Hz, 8 Amps

All Purifiers with model numbers ending in –00 to –09 are designed for operation at 115
volts 60Hz, alternating current. Purifiers with model numbers ending in -10 to -19 are
designed for operation at 100 Volts, 50/60 Hz, alternating current, while –20 to –29 are
designed for operation at 230 +/- 20 volts, 50/60 Hz alternating current. A dedicated
outlet with a circuit breaker rated at 20 amps should be located as close as possible to
the right rear side of the cabinet, at a height even with, or higher than, the bottom of
the cabinet in its final location.
Note: Both electrical outlets of the 115 volt Purifier are protected by a ground fault
interrupter circuit (GFIC). Labconco does not recommend plugging the Purifier into a
GFIC outlet.