Table 5-1

HEPA Filter Leak Test Specifications

Model

34308 34300 34408 34400 34508 34500 34608 34600

Air

Displacement

(CFM) 543 596 723 794 903 991 1083 1189

Laskin

Nozzles

needed 1 1 2 2 2 2 2 2

Calculated aerosol conc.

(ug/l)

1

25 23 38 34 30 27 25 23

Actual aerosol conc. (ug/l)

2

14 12 21 18 17 15 17 15

Nominal Speed Control

Setting

50-60 55-65 65-75 75-85 75-85 80-90 65-75 75-85

1. The calculated concentration was established by the formula:

Concentration (ug/l) = (# Laskin nozzles @ 20 PSIG x 13,500)/Volume of air displaced

2. Based on Mineral oil

1. Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2. Check the oil level of the generator – it must be within 1/8” of the level line

when the generator is sitting level.

3. Establish the 100% and 0% concentration levels for the photometer.

Removing the Exhaust HEPA Filter Cover

On Models Manufactured prior to April 2009 –

4. Remove the exhaust HEPA filter cover by lifting and compressing the cover until

its tabs are clear of the exhaust filter bracket, as shown in Figure 5-2.

Figure 5-2





Lift the center of the cover up while
compressing the cover.




Cover tabs in bracket slots