Figure 7-23
ECM Power Connector Release
Tabs
ECM Data Connector Release Tab
7. Using a 1/2-inch wrench, remove the two upper and one lower blower mounting
bolts, as shown in Figure 7-24.
Figure 7-24
Blower Mounting Bolts
The motor/blower assembly is heavy. Handle with care.
8. While supporting the blower assembly, pull the assembly straight out the
plenum, as shown in Figure 7-25.