Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 156

Advertising
background image

Page 156 of 197

On Models Manufactured After April 2009 –


3. Remove the exhaust HEPA filter cover by loosening the four Phillips Screws on

the sides of the cover, then pivot the front of the cover up, and then pull it
forward as shown.





Pivot the front of the cover up and then pull it
forward (dress panel removed for clarity only).



Loosen these screws only
Note: exercise caution when removing or installing the exhaust filter cover, to prevent
damaging the Exhaust HEPA filter.

4. Locate and remove the four acorn nuts and

lockwashers that secure each exhaust cover
bracket to the top of the Logic. Keep the nuts
and washers; you may discard the brackets.








5. Install the replacement brackets as shown,

with the lockwashers, and tighten the acorn
nuts.


6. Carefully place the canopy between the

brackets. Ensure the canopy fits on the inside
of the exhaust filter flanges assembly as
shown. Line up the holes in the canopy
connection with the screw holes in the filter
flanges.


7. Install the four bolts, washers and

lockwashers supplied with the canopy
connection kit, and tighten them securely to
the exhaust assembly as shown.