PROBLEM

CAUSE

CORRECTIVE ACTION

Fluorescent light
is dim or
flickering

Lamp(s) are
defective

Replace defective lamp(s)

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective lamp
ballast

Replace lamp ballast.

UV light not
working

Sash is open

Close sash – UV light will not work
with the sash open.

Lamp is defective

Replace defective lamp.

Lamp wiring is
disconnected.

Inspect lamp wiring.

Defective

lamp

ballast

Replace lamp ballast.

Keypad

disconnected or
defective

Run keypad diagnostics and check
connections.

UV light is dim or
flickering

Lamp is defective
or is at end of
operating lifetime.

Replace defective or worn out lamp.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective lamp
ballast

Replace lamp ballast.

Airflow Alert
goes off and/or
there is a slight
decrease in filter
life remaining
gauge

HEPA filter loading

The gauge reading steadily
decreases as the Cabinet is used.

Blockage

of

the

return air slots or
grille

Check all return air slots and grilles
to ensure that they are not blocked
or restricted.

Blockage of the

Ensure that the exhaust outlet is