Figure 6-10a & b



The Sensor Distance, as described in Table 6-5,
using the slots cut in the 6-foot template.

4.00-inch

mark

3.25-inch

mark




The Sensor Distance, as described in Table 6-5,
using a scale.






Table 6-5

Secondary Inflow Test Specifications

Model 34410

34610

Restricted Sash Opening

(inches)

2

3 4.25

Sensor distance (inches)

3

4 4

# of Test points

8

12

Test point location

*

*

Nominal Avg. Inflow Vel.

(FPM)

283 231

Avg. Inflow Vel. Range

(FPM)

269-296 220-242

Correction Factor (CF)

1

2

Average Inflow Volume

Range (AIV) (Avg. velocity

x CF)

269-296 403-443

Sash Open Area (Sq. Ft)

2.69

4.03

Inflow Velocity Range
(AIV / Sash open area)

100-110 100-110

1. Locate the single row of holes at the front of the grille. Mark the 6

th

hole from

the side wall and subsequent test points every 9 holes until the number of
test points marked equals the width of the cabinet in feet (for a 4-foot
cabinet, for example, mark the first 4 points). Repeat for the opposite side,
as shown in Figure 6-11.

2. Start the Purifier, and let it operate for at least 5 minutes. Establish the

necessary correction factor to the thermal anemometer to ensure compliance