Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 94
![background image](/manuals/738628/94/background.png)
Page 94 of 197
Figure 6-10a & b
The Sensor Distance, as described in Table 6-5,
using the slots cut in the 6-foot template.
4.00-inch
mark
3.25-inch
mark
The Sensor Distance, as described in Table 6-5,
using a scale.
Table 6-5
Secondary Inflow Test Specifications
Model 34410
34610
Restricted Sash Opening
(inches)
2
3 4.25
Sensor distance (inches)
3
4 4
# of Test points
8
12
Test point location
*
*
Nominal Avg. Inflow Vel.
(FPM)
283 231
Avg. Inflow Vel. Range
(FPM)
269-296 220-242
Correction Factor (CF)
1
2
Average Inflow Volume
Range (AIV) (Avg. velocity
x CF)
269-296 403-443
Sash Open Area (Sq. Ft)
2.69
4.03
Inflow Velocity Range
(AIV / Sash open area)
100-110 100-110
1. Locate the single row of holes at the front of the grille. Mark the 6
th
hole from
the side wall and subsequent test points every 9 holes until the number of
test points marked equals the width of the cabinet in feet (for a 4-foot
cabinet, for example, mark the first 4 points). Repeat for the opposite side,
as shown in Figure 6-11.
2. Start the Purifier, and let it operate for at least 5 minutes. Establish the
necessary correction factor to the thermal anemometer to ensure compliance