Work area air cleanliness test (optional) – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 97

Advertising
background image

Page 97 of 197

Work Area Air Cleanliness Test (optional)

If the customer wants to verify air cleanliness in the work area, use the following
procedure:

1. Mark test points at the geometric center of the work surface, and midway

between the center test point and each side wall.

2. Place a particle counter pickup a test point, at a height of 16 inches above the

work surface, pointing into the airflow, as shown in Figure 6-13. Take three
samples of 1.0 ft

3

. Repeat for each test point.

3. Average the results.

Figure 6-13










Acceptance
The average count should be less than 100 particles 0.5um in size per ft

3

.