Accessing the certifier’s menu, Certification procedures for the type a2 logic, Adjusting the blower speed – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 63

Advertising
background image

Page 63 of 197

Section 5

Certification Procedures for the Type A2 Logic


This section will outline test procedures to validate the performance of the Logic.

Note: These are the only Labconco-approved test procedures for validating
the performance of the Logic Biosafety Cabinet. Other test procedures may
yield different or inaccurate test results.
NOTE: In order to properly certify the Logic, you must obtain a Logic Certifier
Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by
contacting Labconco Customer Service Department at 800-821-5525
.
Note: The ECM motor is programmed by Labconco to deliver a constant
volume of air throughout the life of the HEPA filters. Any changes to the
blower speed control should be small (less than 5 units). If the cabinet does
not balance within 5 units of the model-specific speed control setting shown
in Table 5-1, contact Labconco’s Product Service Department for assistance.

Accessing the Certifier’s Menu

Note: You will need to access this menu to do any of the following:

Adjust the blower speed

To reset the filter gauge

To calibrate the optional airflow sensor

Adjusting the blower speed

1. With the Logic operating, press: “Menu”
2. The Display will change to Menu mode. Press the

or

arrows until “Cabinet

Setup” is selected. Press: “Mute/OK”

3. The Display will then change to the password mode. In sequence, press:

“Light”

“UV Light”

“Timer”

“Timer”

“Mute/OK”

4.

The display will change to the speed setting mode. The number displayed is the
speed control setting, in this case a pulse width modulation (PWM) signal that
indicates the percentage of the motor’s duty cycle. The number should range
from 60 to 90 depending on the Logic model. To increase or decrease the blower
speed, press the

or

buttons, respectively. When the speed control is set

properly, press “Mute/OK” This will set the controller to its new value, and
change the display to the filter gauge mode.