6. Using a 1/2 -inch socket or wrench, loosen the four exhaust filter sealing bolts

located at each corner of the exhaust filter tray, as shown in Figure 7-17, at least
1/2 inch. This releases the exhaust HEPA filter from the cabinet.

Figure 7-17


Exhaust HEPA Filter bolts
(2 per each side)






7. Pull the exhaust HEPA filter straight out of the cabinet.



To Remove the Exhaust HEPA Filter on 3-foot models:

1. Follow steps 1-5 as above.

2. Loosen and remove the four sealed exhaust filter bolts on the top of the cabinet

hull, as shown in Figure 7-18 and 7-19.

3. Remove the exhaust filter pan by lifting it straight up and then out of the

cabinet, as shown in Figure 7-20.

4. When reinstalling the pan, ensure all four edges of the pan are OUTSIDE the

sealing edges of the plenum, as shown in Figure 7-21.

Figure 7-18



Exhaust HEPA Filter bolts