Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 192

Advertising
background image

Page 192 of 197

exhaust outlet

not blocked or restricted.

Blockage

or

restriction under
the work surface

Ensure that the towel catch and
plenum beneath the work surface
are unobstructed.

Contamination of
work in the
cabinet

Improper technique
or procedure for
the Biosafety
Cabinet

See “Use of the Cabinet” section in
the manual.

Restriction

of

the

return air slots or
grille – blockage of
the exhaust outlet

Ensure that all return air slots,
grilles and the exhaust outlet are
unobstructed.

External factors are
disrupting the
cabinet airflow
patterns or acting
as a source of
contamination

See “Installation” section of this
manual.

Cabinet is out of
adjustment/HEPA
filter(s) are
defective

Have cabinet recertified.