Sash Position Switches


Two microswitches are used to indicate the sash position. The switches are the same,
Labconco part # 3832400. A photo of both switches is shown in Figure 8-6.

Figure 8-6



Sash Closure Position Switch











Sash Operating Position Switch



Sash Closure Position Switch

The sash closure position switch signals the display microprocessor that the sash
is open or closed allowing proper operation of the Blower, fluorescent or UV
lights. It is wired in the normally open configuration. If the switch does not
actuate properly when the sash is within ¼-inch of being fully closed, ensure the
switch follower is properly actuating against the weight. See Figure 8-7.

Sash Operating Position Switch


The sash operating position switch signals the display microprocessor that the
sash is open too far for proper operation, as shown in Figure 8-8. It is wired in
the normally open configuration. If the switch does not actuate properly when
the sash is within ¼-inch of its operating position, ensure the switch follower is
properly actuating against the weight.