Exhaust damper adjustment, Sash cable retainer – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 107

Advertising
background image

Page 107 of 197

Sash Cable Retainer

The sash cable retainers are located at the top of each sash pulley, as shown in
Figure 7-9. If the retainer needs adjustment, Use a 3/8-inch wrench to loosen
the bolt securing the retainer and adjust it until it contacts the pulley. Retighten
the bolt.

Figure 7-9









Exhaust Damper Adjustment


The exhaust damper adjustment bolt is located on the left side of the blower plenum
cover, in the center of the orange damper bolt seal. Using a 7/16-inch nut driver or
socket and ratchet, turn the bolt clockwise to open the damper (increase the inflow) or
counterclockwise to close the damper (decrease the inflow). The bolt has 20 turns from
fully open to fully closed. The bolt and adjustment decal is shown in Figure 7-10.

Figure 7-10






Advertising