ADJUSTING THE BLOWER SPEED, FILTER GAUGE
AND AIRFLOW SENSOR

Type A2 with airflow sensor

Press the or keys to adjust the blower
speed. The word ‘wait’ will flash until the
blower speed stabilizes. Press ‘OK’ to
accept the new setting.

CERTIFIER USE ONLY
Calibration mode:
Motor Speed Fac tor:

wait

900

950
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST

80

CERTIFIER USE ONLY
Calibration mode:
Filter setup:
1. New Filter
2. Filter Not Changed
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST

Press the or keys to select. Option # 1 will
reset the filter life remaining gauge to 100%.
Option # 2 will leave the gauge unchanged.
Press ‘OK’ to accept your choice.

The Logic will automatically
turn off and reset itself into
normal operation.

CERTIFIER USE ONLY
Calibration mode:
Downflow:

FPM

UV to autoset
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST

53

CERTIFIER USE ONLY

Calibration mode:
Inflow:

FPM

UV to autoset
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST

105

Press the or keys to adjust the
inflow to match the certified value.
Press ‘UV Light’ to reset the inflow to
105 FPM. Press ‘OK’ to accept.

Press the or keys to adjust the
downflow to match the certified value.
Press ‘UV Light’ to reset the inflow to
55 FPM. Press ‘OK’ to accept.