Sash Cable Retainer
The sash cable retainers are located at the top of each sash pulley, as shown in
Figure 8-9. If the retainer needs adjustment, Use a 3/8-inch wrench to loosen
the bolt securing the retainer and adjust it until it contacts the pulley. Retighten
the bolt.
Figure 8-9
Exhaust Damper Adjustment
The exhaust damper adjustment blade is located on exhaust stack. Using a #2 Phillips
screwdriver, loosen the damper locking screw to move the damper open or closed, as
shown in Figure 8-10.
Figure 8-10
Damper Locking Screw