Hepa filter replacement – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 109

Advertising
background image

Page 109 of 197

6. When the entire seal has been “set”, run a finger around the edge of the seal to

ensure it is properly oriented, as shown in Figure 7-13.

Figure 7-13


HEPA Filter Replacement

CAUTION: The cabinet must be properly decontaminated before servicing the
supply HEPA filter. Following replacement of a HEPA filter, the cabinet must
be recertified.

CAUTION: Always remove the exhaust HEPA filter first. The exhaust filter pan
or plenum can be damaged if the plenum is raised with the Exhaust HEPA
filter in place.

CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate
lifting techniques to remove and handle it. Use two people to remove the
larger filters if possible.

To Remove the Exhaust HEPA Filter on 4-, 5- and 6-foot models:
1. Unplug the cabinet.
2. Remove the front dress panel as shown in Figure 7-1.
3. Loosen all of the Phillips screws on the blower chamber cover. Remove all of the

screws except two screws on the top center of the panel. This will support the panel.

4. Release the damper seal by pushing the outside edge of the seal towards the bolt

and inside the blower plenum cover, as shown in Figures 7-14 and 7-15.