Diffuser removal – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 118
Advertising
![background image](/manuals/738628/118/background.png)
Page 118 of 197
Figure 7-27
Diffuser Removal
Caution: The diffuser should be surface decontaminated before removal.
Note: If an IV bar is in place, it will have to be removed first.
1. Raise the sash to its fully open position.
2. Support the front edge of the diffuser while removing the two 7/16-inch acorn
nuts on the front of the diffuser, as shown in Figure 7-28.
Figure 7-28
Removing the diffuser acorn nuts (Dress
panel and lamps removed for clarity
only).