Electronics module access, Electronic module component identification – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 141
Page 141 of 197
Electronics Module Access
The Electronics module is located on the top right side of the cabinet. To access the
module, do the following:
1. Unplug the Logic, either at the wall, or at the power cord inlet at the back of the
electronics module, as shown in Figure 9-2.
2. Remove the two 7/16-inch acorn nuts and lock washers that secure the module.
3. Tip the module on its side, and secure it so it does not fall.
Figure 9-2
Electronic Module Component Identification
The Electronics module contains the circuit breakers, the relay and power supply boards,
the fluorescent and optional UV ballasts, and the optional RS-232 board. These items
are shown in Figure 9-3.
Figure 9-3
Circuit Breakers
Fluorescent Ballast
UV Ballast
Power Supply Circuit Board (Upper)
Relay Board (Lower)