Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 119
Model 9110TH NOx Analyzer
Setup Menu
Teledyne Analytical Instruments
119
5.9.3.1. Analog Output Voltage / Current Range Selection
In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals. Several other
output ranges are available (see Table 5-6). Each range is usable from -5% to +5% of the rated span.
Table 5-6: Analog Output Voltage Range Min/Max
RANGE NAME
RANGE SPAN
MINIMUM OUTPUT
MAXIMUM OUTPUT
0.1V
0-100 mVDC
-5 mVDC
105 mVDC
1V
0-1 VDC
-0.05 VDC
1.05 VDC
5V
0-5 VDC
-0.25 VDC
5.25 VDC
10V
0-10 VDC
-0.5 VDC
10.5 VDC
The default offset for all VDC ranges is 0-5 VDC.
CURR
0-20 mA
0 mA
20 mA
While these are the physical limits of the current loop modules, typical applications use 2-20 mA or 4-20 mA for the lower and upper
limits. Please specify desired range when ordering this option.
The default offset for all current ranges is 0 mA.
To change the output type and range, select the ANALOG I/O CONFIGURATION submenu (see
Figure 5-5) then press: