Model 9110TH NOx Analyzer

Setup Menu

Teledyne Analytical Instruments

119

5.9.3.1. Analog Output Voltage / Current Range Selection

In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals. Several other
output ranges are available (see Table 5-6). Each range is usable from -5% to +5% of the rated span.

Table 5-6: Analog Output Voltage Range Min/Max

RANGE NAME

RANGE SPAN

MINIMUM OUTPUT

MAXIMUM OUTPUT

0.1V

0-100 mVDC

-5 mVDC

105 mVDC

1V

0-1 VDC

-0.05 VDC

1.05 VDC

5V

0-5 VDC

-0.25 VDC

5.25 VDC

10V

0-10 VDC

-0.5 VDC

10.5 VDC

The default offset for all VDC ranges is 0-5 VDC.

CURR

0-20 mA

0 mA

20 mA

While these are the physical limits of the current loop modules, typical applications use 2-20 mA or 4-20 mA for the lower and upper
limits. Please specify desired range when ordering this option.
The default offset for all current ranges is 0 mA.

To change the output type and range, select the ANALOG I/O CONFIGURATION submenu (see
Figure 5-5) then press: