Display parameter units description – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 89

Advertising
background image

Model 9110TH NOx Analyzer

Overview of Operating Modes

Teledyne Analytical Instruments

89

DISPLAY PARAMETER UNITS

DESCRIPTION

RANGE1
RANGE2

MGM

If AUTO Range mode has been selected, two RANGE functions will
appear, one for each range:

RANGE1: The range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.

RANGE1
RANGE2
RANGE3

If the IND Range mode has been selected, three RANGE functions will
appear, one for each range:

RANGE1:

NO

x

concentration output un A1.

RANGE2: NO concentration output un A2.
RANGE2:

NO

2

concentration output un A3.

NOX STB

STABILITY PPB

The standard deviation of concentration readings of the selected gas.

 Data points are recorded every ten seconds. The calculation uses the

last 25 data points.

SAMP FLW

SAMPFLOW CC/M

Gas flow rate of the sample gas into the reaction cell.

OZONE FL

OZONEFLOW CC/M

Gas flow rate of O

3

gas into the reaction cell.

PMT

PMT MV

The raw signal output of the PMT.

NORM PMT

NORMPMT MV

The signal output of the PMT after is has been normalized for temperature,
pressure, auto-zero offset, but not range.

AZERO

AUTOZERO MV

The PMT signal with zero NO

X

,

which is usually slightly different from 0 V.

This offset is subtracted from the PMT signal and adjusts for variations in
the zero signal.

HVPS

HVPS V

The output power level of the high voltage power supply.

RCELL TEMP

RCELLTEMP C

The temperature of the gas inside the reaction cell temperature.

BOX TEMP

BOXTEMP C

The temperature inside the analyzer chassis.

PMT TEMP

PMTTEMP C

The temperature of the PMT .

IZS TEMP

1

IZSTEMP C

The temperature of the internal span gas generator's permeation tube.

MOLY TEMP

CONVTEMP C

The temperature of the analyzer's

NO

2

NO converter.

RCEL

RCELLPRESS IN-HG-A

The current pressure of the sample gas in the reaction cell as measured at
the vacuum manifold.

SAMP

SAMPPRESS

IN-HG-A

The current pressure of the sample gas as it enters the reaction cell,
measured between the NO/NO

x

and Auto-Zero valves.

NOX SLOPE

NOXSLOPE

The slope calculated during the most recent NO

x

zero/span calibration.

NOX OFFS

NOXOFFSET MV

The offset calculated during the most recent NO

x

zero/span calibration.

NO SLOPE

NOSLOPE

The slope calculated during the most recent NO zero/span calibration.

NO OFFS

NOOFFSET MV

The offset calculated during the most recent NO zero/span calibration.

TEST

TESTCHAN MV

Displays the signal level of the Test Function that is currently being
produced by the Analog Output Channel A4.

TIME

CLOCKTIME HH:MM:SS

The current time. This is used to create a time stamp on DAS readings,
and by the AutoCal feature to trigger calibration events.

1

Only appears if Internal Span Gas Generator option is installed.

To view these TEST functions, press,