Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 69

Advertising
background image

Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

69

Table 3-10:

Valve Operating States OPT 50B installed

MODE VALVE

CONDITION

VALVE PORT

STATUS

SAMPLE

Sample/Cal

Open to SAMPLE inlet

3 2

Zero/Span

Open to ZERO AIR inlet

3 2

Span Shutoff

Closed

Zero Air Shutoff

Closed

ZERO CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to ZERO AIR inlet

3 2

Span Shutoff

OPEN

Zero Air Shutoff

1

Closed

SPAN CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to SPAN inlet

1 2

Span Shutoff

Closed

Zero Air Shutoff

OPEN

3.3.2.5. Zero Scrubber and Internal Span Source (IZS) (OPT 50G)

The 9110T nitrogen oxides analyzer can also be equipped with an internal NO

2

span gas generator and

calibration valve option. This option package is intended for applications where there is a need for
frequent automated calibration checks without access to an external source of span gas.

This valve package includes:

 A 2-stage external scrubber for producing zero air.

 50% Purafil Chemisorbant

®

(for conversion of NO NO

2

).

 50% charcoal (for removal of the NO

2

).

 A heated enclosure for a NO

2

permeation tube.

 This option package DOES NOT contain an actual permeation tube. See

Section 1.4 (Options 52B and 52G) for information on specifying the correct
permeation tube for each application.

 A special desorber that removes all HNO

3

from the calibration gas stream.

 One additional gas inlet port (ZERO AIR).

 One additional gas outlet port (FROM DRYER).

 Two internal valves for switching between the sample gas inlet and the output of the zero/span

subsystem.