Sensor module, reaction cell – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 351

Advertising
background image

Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

351

Input

Ga
in S

elec
to

r 1

11

T

ype K

C

o

m

p

ens

at

ion
4

14

Ou
tpu
t S

ca

le S
ele

cto
r 2 –
1

2

Te
rmi
n

at

ion
S

e

le

ct

or

5

– 1

5

TC2

Not Used

TC1

T

y

p

e J C
o

mp
en
sa

tio
n

3 –
13

Purple Jumpers

Figure 13-19: Thermocouple Configuration Jumper (JP5) Pin-Outs

13.4. SENSOR MODULE, REACTION CELL

The 9110T sensor assembly consists of several subassemblies, each with different tasks:

 The Photo Multiplier Tube (PMT) detects the intensity of the light from the chemiluminescence

reaction between NO and O

3

in the reaction cell. It outputs a current signal that varies in relationship

with the amount of light in the reaction cell.

 The PMT Preamplifier PCA converts the current output by the PMT into a voltage and amplifies it to

a signal strong enough to be usable by the motherboard’s A  D converter.

 It also supplies the drive voltage and gain adjustment for the PMT’s High Voltage Power Supply

(HVPS)

 The Thermo-Electric Cooler (TEC) controls the temperature of the PMT to ensure the accuracy and

stability of the measurements.