Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

347

Table 13-4: Relay PCA Status LED’s

LED Color

Function

Status When Lit

(Energized State)

Status When Unlit

(Default State)

D1

Red Watchdog

Circuit

Cycles ON/OFF every 3 Seconds

under direct control of the analyzer’s CPU.

D2

Yellow

Reaction Cell Heater

Heating

Not Heating

D3 Yellow

NO

2

 NO Converter Heater

Heating

Not Heating

D4

SPARE

D5

1

Yellow

Internal Span Gas Generator

Perm Tube Oven Heater

Heating Not

Heating

D6

SPARE

D7 Green

Zero/Span

Valve Valve OPEN to span gas flow

Valve OPEN to zero gas flow

D8 Green Sample/Cal

Valve

Valve OPEN to

calibration gas flow

Valve OPEN to

sample gas flow

D9

Green

Auto Zero Valve

Sample gas flow BYPASSES

the reaction cell

Sample gas flow is routed

THROUGH the reaction cell

D10 Green

NO/NO

x

Valve

Gas flow routed THROUGH

NO

2

 NO converter

Gas Flow BYPASSES

NO

2

 NO converter

D11

2

Green

Dual Span Gas

Select Valve

Valve OPEN to SPAN 1

gas inlet

Valve OPEN to SPAN2 inlet

D12

3

Green

Pressurized Span

Shutoff Valve

Span gas flow SHUTOFF

Span gas flow OPEN

D13

4

Green

Pressurized Zero

Shutoff Valve

Zero gas flow SHUTOFF

Zero gas flow OPEN

D14 - 16

SPARE

1

Only active when the optional internal span gas generator is installed.

2

Only active when the dual pressurized span option is installed.

3

Only active when one of the pressurized span gas options is installed.

4

Only active when one the pressurized zero gas option is installed.