Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

350

Note

The PMT temperature is maintained by a separate control loop that does
not involve the relay PCA (see Section 13.5.2).

13.3.4.5. Thermocouple Inputs and Configuration Jumper (JP5)

Although the relay PCA supports two thermocouple inputs, the current 9110T analyzers only utilize one.
It is used to sense the temperature of the NO

2

 NO converter.

 This single thermocouple input is plugged into the TC1 input (J15).

 TC2 (J16) is currently not used (see Figure 13-15 for location of J15 and J16).

The type and operating parameters of this thermocouple are set using a jumper plug (JP5).

The default configuration for this thermocouple is:

 Type-K

 Temperature compensated for Type-K

 Isolated

Table 13-5: Thermocouple Configuration Jumper (JP5) Pin-Outs

TC INPUT

JUMPER

PAIR

DESCRIPTION FUNCTION

TC1

1 – 11

Gain Selector

Selects preamp gain factor for J or K TC

OUT = K TC gain factor;

IN = J TC gain factor

2 – 12

Output Scale Selector

Selects preamp gain factor for J or K TC
OUT = 10 mV / °C; IN = 5 mV / °C

3 – 13

Type J Compensation

When present, sets Cold Junction
Compensation for J type Thermocouple

4 – 14

Type K

Compensation

When present, sets Cold Junction
Compensation for K type Thermocouple

5 – 15

Termination

Selector

Selects between Isolated and grounded TC

IN = Isolate TC;

OUT = Grounded TC

TC2

NOT USED


ATTENTION

COULD DAMAGE INSTRUMENT AND VOID WARRANTY

The correct Thermocouple Type must be used if there is ever the need for
replacement. If in doubt please consult Teledyne Customer Service.