Model 9110T NOx Analyzer

Instrument Maintenance

Teledyne Analytical Instruments

240

Table 11-1: 9110T Maintenance Schedule

ITEM ACTION FREQ

CAL

CHECK

REQ’D

DATE PERFORMED

TEST

functions

Review and

evaluate

Weekly No

Particulate

filter

Change

particle filter

Weekly No

Zero/span

check

Evaluate

offset and

slope

Weekly No

Zero/span
calibration

Zero and

span

calibration

Every 3 months

Yes

External zero

air scrubber

option

Exchange

chemical

Every 3 months

No

External

dryer option

Replace

chemical

When indicator

color changes

No

Ozone

cleanser

Change

chemical

Annually

Yes

Reaction cell

window

(“optical

filter” in

Figure 11-6)

Clean Annually

or

as

necessary

Yes

DFU filters

Change

particle filter

Annually No

Pneumatic

sub-system

Check for

leaks in gas

flow paths

Annually or

after repairs

involving

pneumatics

Yes if a

leak is

repaired

Reaction cell

O-rings &

sintered

filters

Replace Annually Yes

PMT Sensor

Hardware

Calibration

Low-level
hardware

calibration

On PMT/

preamp

changes or if

slope is outside

of 1.0±0.3

Yes

Pump

Rebuild head

when RCEL

pressure

exceeds 10 in-

Hg-A (at sea

level)

Yes

Inline

Exhaust

Scrubber

Replace Annually No

NO

2

converter

Replace

converter

Every 3 years

or if conversion

efficiency drops

below 96%

Yes

Desiccant

bags

Replace

Any time PMT

housing is

opened for

maintenance

n/a