Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 293

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

293

12.7.10.2. NO

2

Conversion Efficiency

The efficiency at which the NO

2

 NO converter changes NO

2

into NO directly effects the accuracy of

the 9110T’s NO

x

, NO and NO

2

measurements. The 9110T firmware includes a Converter Efficiency

(CE) gain factor that is multiplied by the NO

2

and NO

X

measurements to calculate the final

concentrations for each. This gain factor is stored in the analyzer’s memory.

The default setting for the NO

2

converter efficiency is 1.0000. Over time, the molybdenum in the NO

2

 NO converter oxidizes and it becomes less efficient at converting NO

2

into NO.

To ensure accurate operation of the 9110T, it is important to check the NO

2

conversion efficiency

periodically and to update this value as necessary.

 For the analyzer to function correctly, the converter efficiency must be between 0.9600 and 1.0200

(96-102% conversion efficiency) as per US-EPA requirements.

 If the converter’s efficiency is outside these limits, the NO

2

converter should be replaced.

 The current converter efficiency level is also recorded along with the calibration data in the DAS for

documentation and performance analysis (Section 7).

12.7.10.3. Calculating / Checking Converter Efficiency

The 9110T to automatically calculate the current NO

2

conversion efficiency by comparing a known

starting concentration of NO

2

gas to the measured NO output of the converter. There are three steps to

performing this operation.

Step 1:

Supply the analyzer with a known concentration of NO

2

gas, to the analyzer.

Figure 12-7:

Setup for determining NO

2

NO Efficiency – 9110T Base Configuration

Step 2:

Advertising