9.6. CALIBRATION QUALITY ANALYSIS

After completing one of the calibration procedures described above, it is important to evaluate the
analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response
curve of the analyzer, separately for NO and NO

X

. The values for these terms, both individually and

relative to each other, indicate the quality of the calibration.

To perform this quality evaluation, you will need to record the values of the following test functions (see
Section 4.1.1), all of which are automatically stored in the DAS channel CALDAT for data analysis,
documentation and archival.

NO OFFS

NO SLOPE

NOX OFFS

NOX SLOPE

Ensure that these parameters are within the limits listed in Table 9-5 and frequently compare them to
those values on the Final Test and Validation Data Sheet (P/N 04490) that came attached to your
manual, which should not be significantly different. If they are, refer to the troubleshooting Section 12.

Table 9-5: Calibration Data Quality Evaluation

Function

Minimum Value

Optimum Value

Maximum Value

NOX SLOPE

-0.700

1.000

1.300

NO SLOPE

-0.700

1.000

1.300

NOX OFFS

-20.0 mV

0.0 mV

150.0 mV

NO OFFS

-20.0 mV

0.0 mV

150.0 mV

The default DAS configuration records all calibration values in channel CALDAT as well as all
calibration check (zero and span) values in its internal memory.

 Up to 200 data points are stored for up 4 years of data (on weekly calibration checks) and a lifetime

history of monthly calibrations.

 Review these data to see if the zero and span responses change over time.

 These channels also store the STB figure (standard deviation of NO

X

concentration) to evaluate if

the analyzer response has properly leveled off during the calibration procedure.

 Finally, the CALDAT channel also stores the converter efficiency for review and documentation.

Advertising