Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 200

Advertising
background image

Model 9110T NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

200

If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which
range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.

9.2.3.1. Setting the Expected Span Gas Concentration

Note

The expected concentrations for both NOx and NO are usually set to the
same value unless the conversion efficiency is not equal to 1.000 or not
entered properly in the conversion efficiency setting.

When setting expected concentration values, consider impurities in your
span gas source (e.g. NO often contains 1-3% NO2 and vice versa).

The NO and NO

x

span gas concentrations should be 80% of range of concentration values likely to be

encountered in your application. The default factory reporting range setting is 500 ppb and the default
span gas concentration is 40.0 ppb.

To set the span gas concentration, press:

Advertising