Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 80
Advertising
Model 9110T NOx Analyzer
Getting Started
Teledyne Analytical Instruments
80
VERIFYING THE EXPECTED NO
X
AND NO SPAN GAS CONCENTRATION
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Verify the PRECISE Concentration Value of the SPAN gases
independently.
If you supply NO gas to the analyzer, the values for expected NO and
NO
x
MUST be identical.
The NO
x
and NO span concentration values automatically defaults to 400.0 PPB and it is recommended
that calibration gases of that concentration be used for the initial calibration of the unit.