Calibration equipment, supplies, and expendables, Spare parts and expendable supplies – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 222

Advertising
background image

Model 9110T NOx Analyzer

EPA Protocol Calibration

Teledyne Analytical Instruments

222

Calibrations should be carried out at the field monitoring site. The analyzer should be in operation for at
least several hours (preferably overnight) before calibration so that it is fully warmed up and its
operation has stabilized. During the calibration, the 9110T:

 Should be in the CAL mode,

 The test atmosphere should be supplied through all components used during normal ambient

sampling and through as much of the ambient air inlet system as is practicable.

 If the instrument will be used on more than one range, it should be calibrated separately on each

applicable range.

 Details of documentation, forms and procedures should be maintained with each analyzer and also

in a central backup file as described in Section 12 of the Quality Assurance Handbook.

 Personnel, equipment and reference materials used in conducting audits must be independent from

those normally used in calibrations and operations.

10.2. CALIBRATION EQUIPMENT, SUPPLIES, AND

EXPENDABLES

The measurement of NO

x

, NO and NO

2

in ambient air requires a certain amount of basic sampling

equipment and supplemental supplies. These include, but are not limited to, the following:

Table 10-1: Activity Matrix for EPA Calibration Equipment and Supplies

Equipment/

Supplies

Acceptance Limits

Frequency And Method

of Measurement

Action If Requirements

Are Not Met

Recorder

Compatible with output signal of

analyzer; min chart width of 150 mm (6

in) is recommended

Check upon receipt

Return equipment to supplier

Sample line and

manifold

Constructed of PTFE, glass or stainless

steel

Check upon receipt

Return equipment to supplier

Calibration

equipment

Meets guidelines and Section 2.3.2 of

Q. A. Handbook

See Section 2.3.9 of Q. A.

Handbook

Return equipment/ supplies to

supplier or take corrective

action

Working standard

NO cylinder gas

Traceable to NIST-SRM SRM. Meets

limits in traceability protocol for

accuracy and stability. Section 2.0.7 of

Q. A. Handbook

Analyzed against NIST-SRM;

see protocol in Section 2.0.7,

Q.A. Handbook

Obtain new working standard

and check for traceability

Recording forms

Develop standard forms

N/A

Revise forms as appropriate

Audit equipment

Cannot be the same as used for

calibration

System must be checked out

against known standards

Locate problem and correct or

return to supplier

When purchasing these materials, a logbook should be maintained as a reference for future procurement
needs and as a basis for future fiscal planning.

10.2.1. SPARE PARTS AND EXPENDABLE SUPPLIES

In addition to the basic equipment described in the Q.A. Handbook, it is necessary to maintain an
inventory of spare parts and expendable supplies. Section 11 describes the parts that require periodic
replacement and the frequency of replacement. Appendix B contains a list of spare parts and kits of
expendables supplies.