Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 267

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

267

TEST FUNCTION

INDICATED FAILURE(S)

concentration incorrect; leaks

NOX OFFset

Incorrect span gas concentration; low-level calibration off

NO SLOPE

HVPS out of range; low-level calibration off; span gas concentration incorrect; leaks

NO OFFSet

Incorrect span gas concentration; low-level calibration off

TIME

Internal clock drifting; move across time zones; daylight savings time?

12.1.3. DIAG SIGNAL I/O: USING THE DIAGNOSTIC SIGNAL I/O

FUNCTION

The signal I/O diagnostic mode allows access to the digital and analog I/O in the analyzer. Some of the
digital signals can be controlled through the touchscreen. These signals, combined with a thorough
understanding of the instrument’s principles of operation (Section 13), are useful for troubleshooting in
three ways:

 The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs and

outputs.

 Many of the components and functions that are normally under algorithmic control of the CPU can

be manually exercised.

 The technician can directly control the signal level Analog and Digital Output signals.

This allows the technician to observe systematically the effect of directly controlling these signals on the
operation of the analyzer. Following is an example of how to use the Signal I/O menu to view the raw
voltage of an input signal or to control the state of an output voltage or control signal.