Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 65

Advertising
background image

Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

65

Figure 3-22:

Pneumatics with Zero/Span Valves OPT 50A

Table 3-9:

Zero/Span Valves Operating States OPT 50A

MODE VALVE

CONDITION

VALVE PORT

STATUS

SAMPLE

Sample/Cal

Open to SAMPLE inlet

3 2

Zero/Span

Open to ZERO AIR inlet

3 2

ZERO CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to ZERO AIR inlet

3 2

SPAN CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to SPAN inlet

1 2